CyBio
®
Well
vario
Release:
2017
109
Fault
Removal
Wrong
parameters
Software
transferred
wrong
parameters
to
sys-
tem.
Check
that
available
soft-
ware
version
is
actually
configured
for
your
system
(refer
to
Hardware system
information
window
in
[Devices]
)
menu.
Check
system
connections
for
proper
state.
On
failure
to
remove
this
error,
you
should
notify
the
customer
service.
Error
on
calculating
motor
characteristic.
Value
settings
for
system
motor
characteristic
are
unrealistic.
Correct
parameter
settings
as
necessary.
Error
on
communicating
with
interface
Interface
transmission
error
which
cannot
be
located
in
greater
detail.
Check
PC
and
system
for
proper
connection.
Check
configuration
of
serial
interface
and
PC.
Error
on
horizontal
drive
motion
Horizontal
motion
was
obstructed
by
incorrectly
seated
microplate.
Remove
microplate
and
mount
it
correctly.
No
residual
ejection
After
aspiration
of
a
vol-
ume,
there
was
no
residual
ejection
of
dosing
stock
during
the
remaining
pro-
gram
sequence.
On
com-
pletion
of
a
program,
there
must
be
no
dosing
liquid
left
in
the
pipette
tips.
Ad
a
[Dispense with
additional stroke]
com-
mand
at
the
appropriate
point
in
a
sequence.
Faulty
horizontal
position
Command
includes
invalid
setting
for
horizontal
posi-
tion.
Check
your
software
and
hardware
configurations
for
mutual
agreement
(refer
to
Hard-
ware system information
window
in
[Devices]
)
menu.
Faulty
height
Command
includes
invalid
height
specification.
Check
your
software
and
hardware
configiurations
for
mutual
agreement.
Error message
Cause
Action for removal
Summary of Contents for CyBio Well vario OL3381-25-300
Page 1: ...Operating Manual CyBio Well vario...
Page 2: ......
Page 4: ......
Page 6: ......
Page 7: ......
Page 8: ......
Page 16: ...XVI Release 2017 CyBio Well vario...
Page 26: ...10 Release 2017 CyBio Well vario Technical Specifications...
Page 60: ...44 Release 2017 CyBio Well vario Technical Description...
Page 74: ...58 Release 2017 CyBio Well vario Initial Start Up Routine Start Up Procedure...
Page 138: ...122 Release 2017 CyBio Well vario Maintenance Care...
Page 144: ...128 Release 2017 CyBio Well vario Accessories Spare Parts...
Page 146: ...130 Release 2017 CyBio Well vario Waste Disposal...
Page 148: ...132 Release 2017 CyBio Well vario...
Page 158: ...142 Release 2017 CyBio Well vario...
Page 160: ...144 Release 2017 CyBio Well vario...
Page 164: ...148 Release 2017 CyBio Well vario...