90
Release:
2017
CyBio
®
Well
vario
Operation
Fig.
94: Insertion of pipetting head (head replacement from the front version)
9.
Tilt
cover
down
until
it
closes,
then
release
cover.
The cover locks by itself. The magazine adapter moves down and un-
locks automatically.
10.
Pull
magazine
adapter
out
of
pipetting
head
from
the
front.
Use
recessed
handle
at
magazine
adapter
bottom
for
this
purpose.
.
Fig.
95: Display – head change – remove magazine adapter
1
Handle
2
Red
marker
3
Rail
4
Recessed
handle
at
adapter
bottom
CAUTION
There
is
danger
of
crushing
or
pinching!
For
opening
or
closing,
hold
the
cover
at
its
handle
with
one
hand!
Keep
other
hand
clear
of
system
space
during
opening
or
closing
motion!
Summary of Contents for CyBio Well vario OL3381-25-300
Page 1: ...Operating Manual CyBio Well vario...
Page 2: ......
Page 4: ......
Page 6: ......
Page 7: ......
Page 8: ......
Page 16: ...XVI Release 2017 CyBio Well vario...
Page 26: ...10 Release 2017 CyBio Well vario Technical Specifications...
Page 60: ...44 Release 2017 CyBio Well vario Technical Description...
Page 74: ...58 Release 2017 CyBio Well vario Initial Start Up Routine Start Up Procedure...
Page 138: ...122 Release 2017 CyBio Well vario Maintenance Care...
Page 144: ...128 Release 2017 CyBio Well vario Accessories Spare Parts...
Page 146: ...130 Release 2017 CyBio Well vario Waste Disposal...
Page 148: ...132 Release 2017 CyBio Well vario...
Page 158: ...142 Release 2017 CyBio Well vario...
Page 160: ...144 Release 2017 CyBio Well vario...
Page 164: ...148 Release 2017 CyBio Well vario...