Operation
40
Edition 06/2015
InnuPure C16
7.
Insert the setup stored in the clipboard via [E
DIT
]
[P
ASTE
].
8.
After copying close the window with the [x] symbol.
9.
Switch the InnuPure C16 off and remove the USB stick.
Editing the setup with a text editor (e.g., Notepad):
1.
Now connect the USB stick to a computer with a text editor installed.
2.
Open the setup in the text editor and empty it as follows:
–
Delete the sample names appearing after "Channel 01: " to "Channel 16: " if en-
tered.
–
Remove the populated equipment type appearing after "Left : " and "Right : ".
Important:
In lines 2 to 19, the colon must be followed by a space! Add a space again if
it was deleted by mistake.
The empty setup should now look as follows (note that the display has been highlighted
to allow the necessary spaces to be recognized more easily):
Figure 11 Empty sample setup in the PC program
3.
Now add the desired sample names at the sample positions to be used (channel 01 to
16), in each case after the last space.
4.
Then enter the corresponding string for the populated equipment for both tray sides (for
information on the assignment of the sides see the chapter "Creating and saving a
sample setup via the HID-Pro 320" p. 37):
Equipment used in the sample tray
String in the sample setup
1 to 4 samples in reagent strips
"Reagent Strips"
1 to 8 samples in a reagent plate
"Reagent Plate"
Empty on one side (without samples) to ensure even
loading of the tray (see the information in the section
"Preparing the sample trays and inserting them into
the InnuPure C16" p. 25)
"None"