Operation
InnuPure C16
Edition 06/2015
33
10. By pressing the [S
TOP
] button, the extrac-
tion protocol can be prematurely aborted
after acknowledging the message "W
OULD
YOU REALLY LIKE TO ABORT
?" with [O
K
].
However, the current command will still be
completed (also read the notes in the sec-
tion "General information and tips for pre-
venting problems" p. 49). Select [C
ANCEL
]
if you want to continue the extraction pro-
tocol. To return to the start window after
aborting a protocol, press the [x] button at
the top right in the window.
Note
During a step in which the filter tips are
ejected the [S
TOP
] button is disabled.
11. After completion of the entire protocol, the
tray remains in the device until it is re-
quested by the user confirming the "Pro-
tocol End" step by means of the [N
EXT
]
button. After the tray moves out of the de-
vice the protocol is automatically closed.
To remove the tray, press lightly against
the tip block. The sample tray is then au-
tomatically moved to the removal position.
CAUTION! Risk of burns!
Do not reach into the device interior after aborting the extraction protocol. The magnetic strip
may be up to 80°C hot!