Operation
36
Edition 06/2015
InnuPure C16
6.3.4
Checking and executing individual device functions with tools
IMPORTANT
Before starting a tool the sample tray has to be pulled in the device or placed in the priming
station. This minimize the risk of a possible error during the automatic pulling in of the sam-
ple tray (see also point 10 in section "General information and tips for preventing problems"
p. 49)
Individual device functions can be executed and checked in the T
OOLS
menu. Using these
tools, the filter tips can for example be emptied or ejected and the tray removed from the
device after aborting an extraction protocol.
1.
In the start window, press the [T
OOLS
]
button.
2.
Select the desired tool (here "
Mag-
net_Test
"
)
and press [S
TART
].
The test run is shown on the screen. Follow
any instructions.
A sample tray must be in the device to execute the tools. If this is not the case when a tool is
started, a prompt to insert a tray is automatically displayed, as already described in the chap-
ter "Starting and executing an extraction protocol" p. 30.