Calibration
Asterion DC Series
3-14
NOTE:
You cannot exit out of Current Ramp Screen using HOME and UP arrow, these
buttons are disabled for the Current Ramp Screen; Refer to Figure 3-25.
Example 1:
Creating a Current ramp using Software Trigger mode
•
Set the
Curr
to 10A
•
Set the
To Curr
to 30A
•
Set the
Volt
to 25V
•
Set the
Time
to 10s
•
Connect an appropriate load to the supply
•
Select the Trigger mode as
SW
(software)
•
Click on
Initialize
•
Click on
Trigger Ramp
•
Observe that
Trigger Ramp
button will change to
Abort
button
•
Observe the current ramp signal using oscilloscope
•
Clicking on the
Abort
button will abort the ramp.
•
Clicking on the
Exit
button will exit the Current Ramp screen.
Example 2:
Creating a Current ramp using Hardware Trigger mode
•
Set the
Curr
to 10A
•
Set the
To Curr
to 30A
•
Set the
Volt
to 25V
•
Set the
Time
to 10s
•
Connect an appropriate load to the supply
•
Select the Trigger mode as
HW
(Hardware)
•
Click on
Initialize
•
Observe that
Trigger Ramp
button will change to
Waiting for Trig
.
•
Give an external trigger i.e. an active high pulse of 10ms on the MOLEX
connector pin-8 (TRIGGER_IN) and pin-12 (DIN_RTN) to generate the
Current Ramp.
•
Observe that
Waiting for Trig
will change to
Abort
button.
•
Observe the current ramp signal using oscilloscope
•
Clicking on the
Abort
button will abort the ramp.
•
Clicking on the
Exit
button will exit the Current Ramp screen.
3.1.7
Configuration Screen
The Configuration power on- settings (PONS), set-up of User V/I Limits, Output Sense,
Measurement Settings and Aux Output.
The top-level menu of the Configuration screen is shown in
. Refer to
Section 3.1.2.1 for navigating to Configuration Screen.
Summary of Contents for Sorensen Asterion DC Series
Page 1: ...M330460 01 Rev D 1 Asterion DC Series Operation Manual...
Page 2: ...Contents Asterion DC Series ii This page intentionally left blank...
Page 23: ...Asterion DC Series Overview M330460 01 Rev D 1 7 Figure 1 2 Auto range Models Characteristics...
Page 38: ...Overview Asterion DC Series 1 22 This page intentionally left blank...
Page 43: ...Asterion DC Series Installation M330460 01 Rev D 2 5 Figure 2 1 Rack mounting 1U Models...
Page 44: ...Installation Asterion DC Series 2 6 Figure 2 2 Rack Mounting 2U Chassis...
Page 49: ...Asterion DC Series Installation M330460 01 Rev D 2 11 Figure 2 5 Rear Panel View 1U Models...
Page 50: ...Installation Asterion DC Series 2 12 Figure 2 6 Rear Panel View 1U Models w EtherCAT Option...
Page 52: ...Installation Asterion DC Series 2 14 Figure 2 8 Rear Panel View 2U Chassis...
Page 83: ...Asterion DC Series Installation M330460 01 Rev D 2 45 This page intentionally left blank...
Page 84: ......
Page 144: ...Calibration Asterion DC Series 3 60 Figure 3 114 Parallel Connection 3 Chassis...
Page 148: ...Calibration Asterion DC Series 3 64 Figure 3 117 Series Connection...
Page 149: ...Asterion DC Series Calibration M330460 01 Rev D 3 1 This page intentionally left blank...
Page 151: ...Asterion DC Series Calibration M330460 01 Rev D 4 3 This page intentionally left blank...
Page 152: ......
Page 157: ...M330460 01 www powerandtest com This page intentionally left blank...