![Ametek Sorensen Asterion DC Series Operation Manual Download Page 32](http://html1.mh-extra.com/html/ametek/sorensen-asterion-dc-series/sorensen-asterion-dc-series_operation-manual_2931304032.webp)
Overview
Asterion DC Series
1-16
5kW and 10kW Auto Range Models
Rated Voltage (V)
Voltage
&Current Rise
Time (ms),
Full load
1
Voltage &
Current Fall
Time (ms),
Full load
3
Voltage Fall
Time (ms), No
load
4
Transient
response
(ms)
5
60
50
80
1000
1
200
50
100
2500
2
400
75
150
3000
2
1)
Maximum time, from 0-100% of programming change from zero to rated output voltage with rated resistive
load.
2)
Maximum time, from 0-100% of programming change from zero to rated output voltage with No load.
3)
Maximum time, from 100%-0 of programming change from rated output voltage to zero with rated resistive
load.
4)
Maximum time, from 100%-0 of programming change from rated output voltage to zero with No load
5)
Typical time to recover within 0.5% of rated output voltage for load step of 10-90% of rated output
current
1.2.11
Remote Sense
Maximum Line
Drop
Compensation
1
40V Model
60V Model
80-600V Models
2V
3V
5V
5V
5V
Connection:
Voltage accuracy/regulation specifications apply at the point where the remote sense leads
are connected.
Line Drop Effect on Output:
Rated output voltage applies at the rear panel output terminals, and line drop
voltage subtracts from the voltage available at the load terminals.
CAUTION!
Due to Line Drop Compensation, if Remote Sense is disconnected from unit
while the output is enabled, output voltage will rise a maximum of 10% of
model’s maximum rated voltage, before faulting
.
1)
Contact factory for additional information regarding higher voltage drop at remote sense operation. Virtual
sense option available uppon request.
Summary of Contents for Sorensen Asterion DC Series
Page 1: ...M330460 01 Rev D 1 Asterion DC Series Operation Manual...
Page 2: ...Contents Asterion DC Series ii This page intentionally left blank...
Page 23: ...Asterion DC Series Overview M330460 01 Rev D 1 7 Figure 1 2 Auto range Models Characteristics...
Page 38: ...Overview Asterion DC Series 1 22 This page intentionally left blank...
Page 43: ...Asterion DC Series Installation M330460 01 Rev D 2 5 Figure 2 1 Rack mounting 1U Models...
Page 44: ...Installation Asterion DC Series 2 6 Figure 2 2 Rack Mounting 2U Chassis...
Page 49: ...Asterion DC Series Installation M330460 01 Rev D 2 11 Figure 2 5 Rear Panel View 1U Models...
Page 50: ...Installation Asterion DC Series 2 12 Figure 2 6 Rear Panel View 1U Models w EtherCAT Option...
Page 52: ...Installation Asterion DC Series 2 14 Figure 2 8 Rear Panel View 2U Chassis...
Page 83: ...Asterion DC Series Installation M330460 01 Rev D 2 45 This page intentionally left blank...
Page 84: ......
Page 144: ...Calibration Asterion DC Series 3 60 Figure 3 114 Parallel Connection 3 Chassis...
Page 148: ...Calibration Asterion DC Series 3 64 Figure 3 117 Series Connection...
Page 149: ...Asterion DC Series Calibration M330460 01 Rev D 3 1 This page intentionally left blank...
Page 151: ...Asterion DC Series Calibration M330460 01 Rev D 4 3 This page intentionally left blank...
Page 152: ......
Page 157: ...M330460 01 www powerandtest com This page intentionally left blank...