Calibration
Asterion DC Series
3-40
voltage programming input. To verify operation in Constant-Voltage mode, follow
these steps:
1. Ensure that there is no load connected to the output.
2. Ensure that the remote sense is connected to the output terminals.
3. Connect a digital voltmeter (DVM) across the rear panel positive and negative
output terminals, observing the correct polarity. Make sure the DVM is in the
DC voltage mode and the range is adequate to handle the full-scale voltage
of the power supply.
4. Apply power to the AC mains input and turn on the power supply.
5. If the Power ON Settings (PONS) had previously been configured to be OFF,
when the supply reaches the Dashboard Screen, enable the output by
pr
essing the “Output On/Off”.
6. Use the Dashboard Screen to program the Voltage, Current and Power.
7. Program the Current to 10% of rated output by entering the value in the
“Setting” section on the Dashboard Screen. Program the current above zero
to enable supplying output current while in the constant-voltage mode.
8.
On the Dashboard screen, rotate the rotary knob to select the “Voltage” text
box in the “Setting” section. Press the rotary knob to highlight the voltage
value. Rotate the rotary knob clockwise and observe both the voltage display
in the “Measure” section on the Dashboard screen and output of the DVM
begin to accelerate up. The output voltage should increase from 0 V to the
maximum rated voltage of the supply. The voltage display in the “Measure”
section on the Dashboard screen and DVM readings should track within the
accuracies of the meter and the Dashboard.
9. Verify the front panel Constant Voltage Mode LED is on.
10. Program the Voltage and Current back to zero.
11. Turn the power supply off.
If Constant-Voltage mode operation did not function as indicated above, verify the
setup and perform the check again. If the function continues to fail, contact the
factory for assistance.
3.2.2
Constant-Current Mode Operation
In Constant-Current mode operation, the output current is regulated at the selected
value while the output voltage varies with the load requirements. The current could
be programmed either through the front panel or by the remote analog current
Summary of Contents for Sorensen Asterion DC Series
Page 1: ...M330460 01 Rev D 1 Asterion DC Series Operation Manual...
Page 2: ...Contents Asterion DC Series ii This page intentionally left blank...
Page 23: ...Asterion DC Series Overview M330460 01 Rev D 1 7 Figure 1 2 Auto range Models Characteristics...
Page 38: ...Overview Asterion DC Series 1 22 This page intentionally left blank...
Page 43: ...Asterion DC Series Installation M330460 01 Rev D 2 5 Figure 2 1 Rack mounting 1U Models...
Page 44: ...Installation Asterion DC Series 2 6 Figure 2 2 Rack Mounting 2U Chassis...
Page 49: ...Asterion DC Series Installation M330460 01 Rev D 2 11 Figure 2 5 Rear Panel View 1U Models...
Page 50: ...Installation Asterion DC Series 2 12 Figure 2 6 Rear Panel View 1U Models w EtherCAT Option...
Page 52: ...Installation Asterion DC Series 2 14 Figure 2 8 Rear Panel View 2U Chassis...
Page 83: ...Asterion DC Series Installation M330460 01 Rev D 2 45 This page intentionally left blank...
Page 84: ......
Page 144: ...Calibration Asterion DC Series 3 60 Figure 3 114 Parallel Connection 3 Chassis...
Page 148: ...Calibration Asterion DC Series 3 64 Figure 3 117 Series Connection...
Page 149: ...Asterion DC Series Calibration M330460 01 Rev D 3 1 This page intentionally left blank...
Page 151: ...Asterion DC Series Calibration M330460 01 Rev D 4 3 This page intentionally left blank...
Page 152: ......
Page 157: ...M330460 01 www powerandtest com This page intentionally left blank...