Asterion DC Series
Calibration
M330460-01 Rev D
3-55
3.6.2
Remote Power Programming by Voltage Source
The DC voltage source is connected between Pin 33 (PPRG_VSOUR) and the
return Pin 18 (RTN_PRG) and Analog Reference source is selected as Voltage from
the front Panel. Refer to Figure 3-108.
The Full-Scale voltage value can be modified to any voltage between 2V to 10V from
front panel screen. Refer to Section 3.1.8.4. Default FSC voltage value is 10V, where
10V corresponds to 100% output power. The corresponding voltage-programming
coefficients for output power are (100% rated output power) / FSC VDC. This
produces transfer functions for output power, as follows:
P
out
= Vdc * (100% rated output power) / 10 VDC), with Vdc in volts, or
P
out
= Vdc * (100% rated output power) / FSC VDC), with Vdc in volts.
Figure 3-108. Remote Power Programming Using 0-10 VDC Source
3.6.3
Remote Power Programming by 4-20mA Source
A 4-20mA current source is connected between Pin 33 (PPRG_VSOUR) and the
return Pin 18 (RTN_PRG) and select the Analog Reference Source as 4-20mA in
front panel. Refer to Figure 3-109.
The transfer function for the output power will be as follows:
P
out
= (I
4-20mA
– 4) (100% rated output power) / 16, with I
4-20mA
in mA
Which sets 0W output power at 4mA and 100% rated power at 20mA.
Figure 3-109. Remote Power Programming Using 4-20mA Source
1
15
16
44
31
30
18
0 2VDC
to
0 10VDC
33 PPRG_VSOUR
1
15
16
44
31
30
18
4 20mA
Current
Source
33 PPRG_VSOUR
Summary of Contents for Sorensen Asterion DC Series
Page 1: ...M330460 01 Rev D 1 Asterion DC Series Operation Manual...
Page 2: ...Contents Asterion DC Series ii This page intentionally left blank...
Page 23: ...Asterion DC Series Overview M330460 01 Rev D 1 7 Figure 1 2 Auto range Models Characteristics...
Page 38: ...Overview Asterion DC Series 1 22 This page intentionally left blank...
Page 43: ...Asterion DC Series Installation M330460 01 Rev D 2 5 Figure 2 1 Rack mounting 1U Models...
Page 44: ...Installation Asterion DC Series 2 6 Figure 2 2 Rack Mounting 2U Chassis...
Page 49: ...Asterion DC Series Installation M330460 01 Rev D 2 11 Figure 2 5 Rear Panel View 1U Models...
Page 50: ...Installation Asterion DC Series 2 12 Figure 2 6 Rear Panel View 1U Models w EtherCAT Option...
Page 52: ...Installation Asterion DC Series 2 14 Figure 2 8 Rear Panel View 2U Chassis...
Page 83: ...Asterion DC Series Installation M330460 01 Rev D 2 45 This page intentionally left blank...
Page 84: ......
Page 144: ...Calibration Asterion DC Series 3 60 Figure 3 114 Parallel Connection 3 Chassis...
Page 148: ...Calibration Asterion DC Series 3 64 Figure 3 117 Series Connection...
Page 149: ...Asterion DC Series Calibration M330460 01 Rev D 3 1 This page intentionally left blank...
Page 151: ...Asterion DC Series Calibration M330460 01 Rev D 4 3 This page intentionally left blank...
Page 152: ......
Page 157: ...M330460 01 www powerandtest com This page intentionally left blank...