AMETEK CTS
OCS 500 N6 series
Manual for Operation
V 5.24
64 / 80
10.3. Verification Ringwave
10.3.1. Verification Parameter
IEC 61000-4-12, ANSI C62.41
(open circuit)
V peak open circuit voltage
±10%
Rise time first peak
0.5 μs
±20%
Oscillating frequency
100 kHz
±10%
Decaying of peak 2-1 40% to 110%
Decaying of peak 3-2 40% to 80%
Decaying of peak 4-3 40% to 80%
(Short circuit)
I peak (12
)
333.33A
±10%
I peak (30
)
133.33A
±10%
Rise time first peak
0.2 to 1.0
s
Oscillating frequency
100 kHz
±10%
Impedance
12
/ 30
±20%
Remark:
The output impedance is calculated result of the ratio of the first peak open circuit voltage and the
first peak short current, measured with a 4000V pulse.
Typical pulses, measured with a calibrated equipment
Ringwave coupling L1
– N impedance = 30
Open
circuit
Short
circuit
U set : +4000V
U set : -4000V
Ringwave coupling L1
– N impedance = 12
Short
circuit
U set : +4000V
U set : -4000V