AMETEK CTS
OCS 500 N6 series
Manual for Operation
V 5.24
29 / 80
5.3.
RINGWAVE Immunity requirements as per ANSI C 62.41 (IEC 61000-4-12)
Test level
Open circuit voltage
250 V
– 6‘000 V ± 10 %
Wave shape
open circuit
Rise time first peak T1
0.5
μs ± 30 %
Oscillation frequency 1/T
100 kHz ± 10 %
Decaying of Pk1 to Pk2
40 % - 110 %
Decaying of Pk2 to Pk3
40 % - 80 %
Decaying of Pk3 to Pk4
40 % - 80 %
Decaying other peaks
no requirements for other peaks
Output impedance
12
Ω, 30 Ω ± 20 % (internal depend setting)
12
Ω internal ± 20 % (with setting R external)
Wave shape
short circuit
Rise time first peak tr T1
0.2 to 1.0
s (short circuit)
Oscillation frequency 1/T
100 kHz ± 10 %
Decaying
no requirements
Short circuit current 12Ω
500 A ± 10 %
Short circuit current 30Ω
200 A ± 10 %
(+10 % - 0 %, models as per IEC 61008-1 for RCCB)
Polarity
positive / negative / alternating
Repetition rate
250 V
– 4000 V 1 s – 999 s
> 4000V
10 s
– 999 s
Events preselection
1 - 30'000 or endless
Counter
1
– 100’0000
Trigger
Trigger of pulses
AUTO, MAN, EXTERN
Synchronization
0 - 360° (16
– 500 Hz)
Asynchron = 0° If a reference signal is connected to the Sync input.
Resolution
1°
Output
Direct
HV-COM connector Zi = 12
Ω or 30 Ω depend setting
Coupling network 1 phase
To L, N, PE all combinations
Coupling network 3 phase
L1...L3-N, L1...L3-PE, N-PE, L1+L2+L3-PE , L1+L2+L3+N-PE
Measurements
CRO
5V Trigger
Peak voltmeter
6000 V ± 10 % 50 V resolution
Peak current meter
500 A ± 10%
5 A resolution
Test routines
Quick Start
Immediate start, all parameters adjustable during a running test
Standard test routines as per
IEC 61000-4-12 level 1
IEC 61000-4-12 level 2
IEC 61000-4-12 level 3
IEC 61000-4-12 level 4
IEC 61000-4-12 Manual operated standard routine (single phase equipment)
User test routines
Change polarity after n pulses
Change voltage level V after n pulses by
V
Change phase angle A after n pulses by
A
Change coupling mode after n pulses
Summary of Contents for OCS 500 M6
Page 76: ......