AMETEK CTS
OCS 500 N6 series
Manual for Operation
V 5.24
51 / 80
8.1.1.
Quick Start
Easy and very fast operation of all standard functions of the equipment. The latest simulator settings are stored
automatically and will be recalled when Quick Start is next selected.
Page 2
Ringwave
Quick start
V
= 2000
V
Imp =
12 Ohm
A
= 0 dgr
+/-
=
+
cpl
= L
–PE
tr
=
1
s
tri
= Auto
n
=
1001 pulses
START CHANGE
F1
F2
F3
F4
F5
F6
F7
Press
START
and the test routines begin to work.
Press
CHANGE
and the actual parameters can be changed.
All function keys except F2 (manual trigger) can
Stop
the test routine.
Page 3
(Start)
Ringwave
Quick start
V
=
2000V
Imp =
12 Ohm
A
= 0 dgr
+/-
=
+
cpl
= L
–PE
tr
=
1
s
tri
= Auto
n
=
1001 pulses
VSet =
2000V
U = + 2000V
COUNTER
STOP
5s
I = + 165A
43
F1
F2
F3
F4
F5
F6
F7
The user can select the parameter with the related function key and can change the value with the knob. The
cursor allows the user to define the value of the digit which should be changed (fast or slow change).
Pressing the ESC button will bring the user back to the previous level from where the test can be restarted with
new levels. After restart the actual test time is displayed. All functions keys except F2 (MAN TRIGGER) can stop
the test routine. The latest setting will be displayed.
Any pressing of a function key will indicate the functions START, CHANGE or CONTINUE. F3 will continue the
same test routine. Also the test time will continue running. If the user selects at first START or CHANGE the test
will be stopped completely.
Page 3
(Change)
Ringwave
Quick start
Ringwave
Quick start
Voltage
V
: 250V-
6000V
Repetition tr
: 1s
- 999s
Impedance Imp : 12
30
Trigger
tri
: Auto / Man
Angle
A
: 0grd
- 360grd / asyn.
Events
n
: 1
- 30000 / Endless
Coupling cpl
: L / N / PE
U Imp A +/- cpl tr
tri n
2 / 2
250 12 Ohm 0 + L-PE 2 1 / 2
Auto 100
F1
F2
F3
F4
F5
F6
F7
F1
F2
F3
F4
F5
F6
F7