8.1.3.
User Test Routines
The user can program, save and recall his own specific test routines. The next pages show the selection of the
functions.
Ringwave
IEC 61000-4-12
F1 : Change polarity after n pulses
F2 : Voltage change after n pulses by
V
F3 : Change Angle after n pulses by
a
F4 : Change coupling after n pulses
F1
F2
F3
F4
F5
F6
F7
After selection the last used test parameters will be indicated on the display.
Customized test routines
The software controls user test routines according to the specification of the user. All limitations are the same as
defined under Quick Start.
F1 Change polarity after n pulses
After the release of the preselected number of pulses the
polarity is changed. The procedure always starts with
positive polarity and changes than to negative. The same
parameters as under Quick Start can be selected.
F2 Voltage change after n pulses by
V
The test voltage V is changed from V1 to V2. After the preselected
number pulses the test level is changed by
V until V2 is reached.
The same parameters as under Quick Start are selectable.
For the limitation of the max. Admissible repetition rate the
higher value of V1 and V2 is valid.
F3 Change the phase angle after n pulses by
A
The phase angle related to which the surge pulse is
released is changed from A1 to A2. After the preselected
number of n pulses the actual phase angle is changed by
A
until A2 is reached. The same parameters as under Quick
Start can be selected.
F4 Change coupling after n pulses
The coupling mode will automatically be changed after the
preselected number of pulses has been released. All
possible coupling modes will be selected. The same
parameters as under Quick Start can be selected.