AMETEK CTS
OCS 500 N6 series
Manual for Operation
V 5.24
52 / 80
8.1.2.
Standard test Routine
Page 3
Ringwave
IEC 61000-4-12
F1 : Common Mode
F2 : Differential Mode
F1
F2
F3
F4
F5
F6
F7
Page 4
Ringwave
IEC 61000-4-12
With the selection of F1 ....F4 the test is conducted
automatically with the parameters and the test sequence
as required per IEC 61000-4-12. The only parameters
the operator is able to change are the
-
impedance
-
repetition rate
-
number of pulses per test level
- trigger Auto / Manual
F1 : Level 1
500 V
F2 : Level 2
1000 V
F3 : Level 3
2000 V
F4 : Level 3
4000 V
F7 : Manual
F1
F2
F3
F4
F5
F6
F7
8.1.2.1.
Level test
A selected level will step through all couplings in the selected Mode. The following parameters are used:
Defined per
standard
Voltage
selected level voltage
Polarity
each coupling
positive and negative
Coupling
Differential Mode
L-N
Common Mode
L-PE, N-PE, L+N-PE
Variable
parameters
Impedance: imp
12
or 30
Repetition:
tr
0.0 s
– 999 s
trigger:
tri
Auto, Manual
Events:
N
1 - 30000
Ringwave
IEC 61000-4-12 Level 3
Ringwave
IEC 61000-4-12 Level 3
2000V
2s
AUTO
2000V
2s
AUTO
5 pulses
12 Ohm
5 pulses
12 Ohm
- 1000 V
L
L-N
90 deg
Counter
V = -980 V
act coupling
4
START CHANGE
STOP Step
I = - 85.A
complete
29
F1
F2
F3
F4
F5
F6
F7
F1
F2
F3
F4
F5
F6
F7
- The counter shows the number of triggered pulses per actual coupling as well as the number of all triggered test
pulses within the running test sequence.
- Pushing the function key F2 STEP will bring you into the next iteration sequence.