AMETEK CTS
OCS 500 N6 series
Manual for Operation
V 5.24
61 / 80
9.2.
Menu Damped Osc. Wave 1 MHz magnetic field (MF)
There are the same menu functions for the 100 kHz and 1 MHz damped oscillating magnetic field.
Page 2
(Select function)
Damped Osc. 1MHz MF
IEC 61000-4-10
F1 : Quick Start
F2 : Standard test routines
F3 : User test routines
F7 : Magnetic field correction factor
F1
F2
F3
F4
F5
F6
F7
F7 Magnetic field correction factors
The operator may enter:
F1: Antenna factor
Af
1/m
= 0.90 (MS100N)
For a magnetic field of
30Am
-1
the current in the Antenna is 30 Am
-1
/ 0.90 m
-1
=
33.3A
20%
These values are given together with the delivered antenna. The antenna factor depends on the type of antenna
which is used for the test. The operation of the pulsed oscillating magnetic field test is similar as to the standard
oscillating routines. For magnetic field testing the antenna correction factor shall be entered for fine tuning the
current. The operator can enter this factor within the setup menu under the service routine.
Page 2
(show parameter)
Damped Osc. 1MHz MF
Quickstart
H max for magnetic field test = 100A/m
Recommended antenna factor for 100Am
-1
:
Frequency = 100kHz :
0.90
Frequency =
1 MHz :
0.90
H
= 30A/mA
Rep =
400/s
td
=
2s
tr
=
10s
cop
= /
+/-
=
+
T
= 00:10 min
START CHANGE
F1
F2
F3
F4
F5
F6
F7
Press
START
and the test routines begin to work.
Press
CHANGE
and the actual parameters can be changed.
Setup of pulsed magnetic Test field
At 1 MHz test, the cables (max. 3 m) must be
twisted
Connect the antenna with two banana cables to the HV and COM plug on the OCS 500 N6x model
The release of the damped oscillated impulse is mostly related to a certain phase angle. The impulses are
synchronized to the input signal at the rear Sync-connector.
Warning
High Voltage
Danger of life by touching the antenna!
Summary of Contents for OCS 500 M6
Page 76: ......