AMETEK CTS
OCS 500 N6 series
Manual for Operation
V 5.24
46 / 80
7.1.2.
Standard test routines
The user can select preprogrammed standard test routines.
Page 3
Damped Osc. 3MHz
IEC 61000-4-18
F1 : Common Mode
F1
F2
F3
F4
F5
F6
F7
Page 4
Damped Osc. 3MHz
IEC 61000-4-18
F1 : Level 1
500 V
F2 : Level 2
1000 V
F3 : Level 3
2000 V
F4 : Level 4
4000 V
F7 : Manual
F1
F2
F3
F4
F5
F6
F7
The standard levels for common and differential mode can be changed in the setup menu.
7.1.2.1.
Level test
A selected level will step through all couplings in the selected Mode. The following parameters are used:
Defined per
standard
Voltage
selected level voltage
Repetition f
5000/s (3MHz)
5000/s (10 MHz)
5000/s (30 MHz)
Polarity
each coupling
positive and negative
Coupling
Common Mode
Lx-PE, N-PE, Lx+N-PE
Variable
parameters
Burst Time td
1 ms
– 50 ms
Depends on fast module setting 3, 10, 30 MHz
Repetition tr
0.0 s
– 99.9 s
Test duration
1 min
Damped
Page 5
(standard test in stop mode)
(Running standard test)
Damped Osc. 3 MHz
IEC 61000-4-18 L3
Damped Osc. 3 MHz
IEC 61000-4-18 L 3
V
=
2000 V
Rep
= 5000/s
V
=
2000 V
Rep
= 5000/s
Td
= 50ms
Td
= 50ms
Tr
= 0.3s
+ 2000 V
Tr
= 0.3s
T
= 1:00min
L
Testtime
current coupling
0:00:21
START CHANGE
STOP Step
complete test time
0:02:21
F1
F2
F3
F4
F5
F6
F7
F1
F2
F3
F4
F5
F6
F7
- The Test time shows the elapsed time per actual test level as
well as the complete time of all triggered test levels within the
running test sequence.
- Pushing the function key F2 STEP will bring you into the next
iteration sequence.