MRO Output Water Quality
The physician in charge of dialysis has the ultimate responsibility for selecting the maximum
allowable levels of chemical contaminants in the water, and, also, is responsible for
monitoring the water. The AmeriWater MRO System is designed to produce water that
meets or exceeds ANSI/AAMI RD62 requirements.
Thin Film Composite Membrane
Contaminants
Percentage Removal
Calcium
Magnesium
Sodium
Potassium
Fluoride
Nitrate (NO
3
)
Sulfate
Copper
Barium
Zinc
Aluminum
Arsenic
Lead
Silver
Cadmium
Chromium
Selenium
Mercury
Antimony
Beryllium
Thallium
99.5
99.5
98.0
97.0
87.0 - 93.0
60.0 - 75.0
99.5
98.0 - 99.0
96.0 - 98.0
98.0 - 99.0
98.0 - 99.0
94.0 - 96.0
96.0 - 98.0
93.0 - 96.0
96.0 - 98.0
96.0 - 98.0
94.0 - 96.0
96.0 - 98.0
96.0 - 98.0
96.0 - 98.0
96.0 - 98.0
PT401 Antiscalant/Scale Inhibitor
Contaminants
Chemical Feed
System
Membrane Scale Control
Not to exceed 40 ppm
2.3
TEMPERATURE CORRECTED MRO PRODUCTION RATES
MRO membrane performance is affected by water temperature. The Product Water Flow
Rate and Output decreases as the temperature of the Incoming Tap Water decreases.
Summary of Contents for MRO3
Page 8: ...MRO 3 4 MANUAL 5 98 0140 Rev F SECTION 3 COMPONENTS AND SCHEMATICS 3 1 EXTERNAL FRONT VIEW...
Page 10: ...MRO 3 4 MANUAL 7 98 0140 Rev F 3 2 INTERNAL REAR VIEW...
Page 12: ...MRO 3 4 MANUAL 9 98 0140 Rev F 3 3 ELECTRICAL DIAGRAM MRO3 MRO4...
Page 13: ...MRO 3 4 MANUAL 10 98 0140 Rev F 3 4 FLUID DIAGRAM MRO3 MRO4...
Page 34: ...MRO 3 4 MANUAL 31 98 0140 Rev F FIGURE 6 2...
Page 35: ...MRO 3 4 MANUAL 32 98 0140 Rev F FIGURE 6 3...
Page 52: ...MRO 3 4 MANUAL 49 98 0140 Rev F SECTION 11 MRO3 4 SPARE PARTS LISTING...