MRO 3 & 4 MANUAL
3
98-0140
Rev. F
SECTION 2 TECHNICAL INFORMATION
2.1
SPECIFICATIONS
Ideal, minimum, and maximum
incoming water temperature
Min = 41° F (5° C) Max = 90° F (33° C)
Ideal Temperature = 77° F (25° C)
Inlet gauge pressure (when the
MRO is running)
Minimum
Maximum
20 PSI
90 PSI (
P
ounds per
S
quare
I
nch)
Pump pressure –
Minimum
Maximum
160 PSI
270 PSI
Water pressure to dialysis machine
Controlled by Loop return relief valve.
Maximum output of product water
@ 77°F (25°C), TDS<1000 ppm of
NaCl, & pump pressure of 150 psi.
MRO3 - 4200 GPD (15,897 LPD)
MRO4 - 5600 GPD (21,196 LPD)
(
G
allons
P
er
D
ay /
L
iters
P
er
D
ay)
Connections
Feed =3/4” plain hose
Product = 1/2" plain hose
Drain = 1/2" plain hose
Electrical Requirements
For the controller: 115V/60Hz GFI (
G
round
F
ault
I
nterrupter)
For the pump: 230V single phase/15 Amp; or
208V, 230V, or 460V three phase/10 Amp fused
disconnect
Dimensions Packaged
Dimensions not Packaged
58" H x 42" W x 38" D
52”H x 18 ½”W x 18 ½”D
Shipping Weight
275 - 300 lbs.
Materials that Contact Product Water:
ABS
Polyethylene
Acrylic
Polypropylene
Carbon
Stainless Steel
Nylon
Thin Film Composite Membrane (polyimide)
PVC
Tygon
Polyester
All of the above listed materials meet FDA and/or NSF standards.
Summary of Contents for MRO3
Page 8: ...MRO 3 4 MANUAL 5 98 0140 Rev F SECTION 3 COMPONENTS AND SCHEMATICS 3 1 EXTERNAL FRONT VIEW...
Page 10: ...MRO 3 4 MANUAL 7 98 0140 Rev F 3 2 INTERNAL REAR VIEW...
Page 12: ...MRO 3 4 MANUAL 9 98 0140 Rev F 3 3 ELECTRICAL DIAGRAM MRO3 MRO4...
Page 13: ...MRO 3 4 MANUAL 10 98 0140 Rev F 3 4 FLUID DIAGRAM MRO3 MRO4...
Page 34: ...MRO 3 4 MANUAL 31 98 0140 Rev F FIGURE 6 2...
Page 35: ...MRO 3 4 MANUAL 32 98 0140 Rev F FIGURE 6 3...
Page 52: ...MRO 3 4 MANUAL 49 98 0140 Rev F SECTION 11 MRO3 4 SPARE PARTS LISTING...