MRO 3 & 4 MANUAL
17
98-0140
Rev. F
SECTION 5 DISINFECTING THE SYSTEM
5.1
DISINFECTING THE SYSTEM
The MRO system should be disinfected according to specifications of your medical facility
director. As a general guideline, AmeriWater recommends that the system should be
disinfected at least monthly. In addition, AmeriWater recommends that the system be
disinfected if it has not been: used for 24 hours or flushed at least every 8 hours.
5.1.a DISINFECTING THE MRO ONLY OR WITH A SMALL LOOP SYSTEM
Note: A “small loop system” can be defined as one where all points in the loop can
easily be disinfected by drawing less than 5 containers of PAA solution into the MRO
and loop system. If this can’t be accomplished, refer to section 5.1.b that describes
disinfecting a “large” loop system.
1.
Switch off the MRO by pressing the POWER key (the display will show STANDBY).
2.
Turn off the feed water supply to the MRO.
3.
Place a container under the
CHLORAMINES SAMPLE PORT to relieve the pressure on the system by slowly
opening the Sample Port.
4.
Disconnect any dialysis machines from the Direct Feed Loop or Central System from
the water loop before starting the disinfection procedure. Disinfectant for the MRO or
loop must never be allowed to enter a dialysis machine.
FIGURE 5.1
Summary of Contents for MRO3
Page 8: ...MRO 3 4 MANUAL 5 98 0140 Rev F SECTION 3 COMPONENTS AND SCHEMATICS 3 1 EXTERNAL FRONT VIEW...
Page 10: ...MRO 3 4 MANUAL 7 98 0140 Rev F 3 2 INTERNAL REAR VIEW...
Page 12: ...MRO 3 4 MANUAL 9 98 0140 Rev F 3 3 ELECTRICAL DIAGRAM MRO3 MRO4...
Page 13: ...MRO 3 4 MANUAL 10 98 0140 Rev F 3 4 FLUID DIAGRAM MRO3 MRO4...
Page 34: ...MRO 3 4 MANUAL 31 98 0140 Rev F FIGURE 6 2...
Page 35: ...MRO 3 4 MANUAL 32 98 0140 Rev F FIGURE 6 3...
Page 52: ...MRO 3 4 MANUAL 49 98 0140 Rev F SECTION 11 MRO3 4 SPARE PARTS LISTING...