MRO 3 & 4 MANUAL
40
98-0140
Rev. F
3.
Turn on the MRO and allow water to run through the system for a minimum of 2 hours.
This will rinse the preservative out of the new membrane.
CAUTION:
The membrane is not rinsed thoroughly until the water is clear!
4.
Turn the knob on the Reject Flow Meter clockwise so that the Product Water flow and
Reject Water flow are approximately equal, then allow the MRO to run. Run the MRO
about 30 minutes after the conductivity is below the setpoint, and no longer alarming.
Turn off the MRO.
5. Disinfect the system per the instructions in section 5.1.
6.
Pull an AAMI analysis. The MRO can be used after disinfection when it begins
producing water to the specifications.
7. Reconnect the PRODUCT WATER hose to the dialysis machine.
8. Turn on the MRO. The rinse out cycle is now complete, and the MRO is ready for use.
WARNING:
If the product water conductivity does not come out of alarm, do not use
the system! Continue rinsing, or call AmeriWater for guidance.
8.6
AMERIWATER CLEAN IN PLACE (CIP)
1. Move the CLEAN IN PLACE SWITCH (inside the MRO on
the back of the controller) to the ON position.
2. Follow the CIP instructions for connecting the drum to the
MRO.
3. Remove the carbon tank and install the provided jumper.
4. Follow the instructions provided with the CIP system.
Summary of Contents for MRO3
Page 8: ...MRO 3 4 MANUAL 5 98 0140 Rev F SECTION 3 COMPONENTS AND SCHEMATICS 3 1 EXTERNAL FRONT VIEW...
Page 10: ...MRO 3 4 MANUAL 7 98 0140 Rev F 3 2 INTERNAL REAR VIEW...
Page 12: ...MRO 3 4 MANUAL 9 98 0140 Rev F 3 3 ELECTRICAL DIAGRAM MRO3 MRO4...
Page 13: ...MRO 3 4 MANUAL 10 98 0140 Rev F 3 4 FLUID DIAGRAM MRO3 MRO4...
Page 34: ...MRO 3 4 MANUAL 31 98 0140 Rev F FIGURE 6 2...
Page 35: ...MRO 3 4 MANUAL 32 98 0140 Rev F FIGURE 6 3...
Page 52: ...MRO 3 4 MANUAL 49 98 0140 Rev F SECTION 11 MRO3 4 SPARE PARTS LISTING...