158
R
EV
9
A
MERICAN
M
AGNETICS
, I
NC
.
R
EMOTE
I
NTERFACE
R
EFERENCE
: C
OMMAND
R
EFERENCE
manufacturer as the maximum safe limit of magnet operation (for AMI
magnets it is the Magnet Rated Current, see page 108).
•
CURRent:LIMit?
Returns the Current Limit in amperes.
•
CONFigure:QUench:DETect {0|
1
|2|3}
Sending “0” disables the automatic quench detection function of the
Model 430 Programmer. “1” enables the current-mismatch quench
detection function of the Model 430. “2” enables the temperature-limit
quench detection function. “3” enables both the current-mismatch and
temperature-limit methods of quench detection with either able to trigger
a quench. See page 123 for more information. “1” is the default value.
•
QUench:DETect?
Returns “0” indicating automatic quench detection is disabled, or “1”
indicating that the current-mismatch automatic quench detection is
enabled. “2” indicates temperature-limit quench detection is enabled. “3”
indicates
both
current-mismatch and temperature-limit quench detection
methods are enabled.
•
CONFigure:QUench:RATE {1|2|
3
|4|5}
Sets the value of the quench detect sensitivity. The range is from 1 to 5
in integer increments. “1” indicates the most sensitivity. “5” is the least
sensitivity. “3” (normal) is the default value.
•
QUench:RATE?
Returns the value of the quench detect sensitivity from 1 to 5 in integer
increments where “1” is the most sensitivity and “5” is the least.
•
OPLimit:IC?
Returns the present value of the temperature-based maximum (i.e.
critical) current limit,
Ic
, specified by the equation documented on
•
OPLimit:TEMP?
Returns the temperature in units of Kelvin read from the Auxiliary Input 3
(see page 202).
•
CONFigure:OPLimit:MODE {
0
|1|2}
Sets the Protection Mode for the Operational Limits submenu (refer to
section on page 91). Sending “0” sets the mode is Off. Sending “1” sets
Summary of Contents for 430
Page 2: ......
Page 16: ...XVI REV 9 AMERICAN MAGNETICS INC FOREWORD SAFETY SUMMARY ...
Page 28: ...12 REV 9 AMERICAN MAGNETICS INC INTRODUCTION OPERATING CHARACTERISTICS ...
Page 64: ...48 REV 9 AMERICAN MAGNETICS INC INSTALLATION POWER UP AND TEST PROCEDURE ...
Page 208: ...192 REV 9 AMERICAN MAGNETICS INC SERVICE RETURN AUTHORIZATION ...
Page 248: ...232 REV 9 AMERICAN MAGNETICS INC APPENDIX SHORT SAMPLE MODE ...