COILconst?
Returns the coil constant setting in kG/A or T/A per the selected field
units.
•
CONFigure:INDuctance <
inductance (H)
>
Sets the currently-connected magnet inductance in Henries.
•
INDuctance:SENSe?
Initiates a direct measurement of the currently connected magnet. The
measurement requires the magnet to be ramping up or down while not
passing through a ramp segment change during the measurement. If
these conditions are not met, the Model 430 will beep once, generate an
error, and exit the measurement. Otherwise, this will return a
measurement of the magnet inductance in Henries. The measured value
will automatically be stored as the currently-connected inductance.
NOTE
The inductance measurement
requires
the magnet voltage input
to the rear panel Magnet Station connector as described on
page 193. Please note that some magnet system manufacturers
do not provide magnet voltage taps.
NOTE
This remote command blocks communication for several seconds
on all ports until the measurement process completes or is
interrupted.
•
INDuctance?
Returns the currently-connected magnet inductance in Henries.
•
CONFigure:ABsorber {
0
|1}
Sending “0” indicates that an energy absorber is not present in the
system. A “1” indicates that an energy absorber is present. “0” is the
default value.
•
ABsorber?
Returns “0” indicating that an energy absorber is not present in the
system, or “1” indicating that an energy absorber is present.
Summary of Contents for 430
Page 2: ......
Page 16: ...XVI REV 9 AMERICAN MAGNETICS INC FOREWORD SAFETY SUMMARY ...
Page 28: ...12 REV 9 AMERICAN MAGNETICS INC INTRODUCTION OPERATING CHARACTERISTICS ...
Page 64: ...48 REV 9 AMERICAN MAGNETICS INC INSTALLATION POWER UP AND TEST PROCEDURE ...
Page 208: ...192 REV 9 AMERICAN MAGNETICS INC SERVICE RETURN AUTHORIZATION ...
Page 248: ...232 REV 9 AMERICAN MAGNETICS INC APPENDIX SHORT SAMPLE MODE ...