CONFigure:LOCK:RAMPDown {
0
|1}
Specifies whether the External Rampdown Enabled picklist value (under
the Protection Setup submenu) is locked or unlocked. Sending “0”
unlocks. A “1” locks. “0” is the default value.
•
LOCK:RAMPDown?
Returns “0” for External Rampdown Enabled picklist value (under the
Protection Setup submenu) unlocked, or “1” for locked.
N
ET
S
ETUP
C
ONFIGURATION
C
OMMANDS
AND
Q
UERIES
The Net Setup Configuration Commands and Queries provide read/write
access to the
IPNAME
setting available in the Net Setup submenu
•
CONFigure:IPNAME <
system name
>
Sets the system name (also known as
host name
or
computer name
),
the
name
by which the Model 430 Programmer is identified on a
network.
NOTE
If the system name value is changed, the Model 430 Programmer
power must be cycled off for at least 5 seconds and then back on
to complete the change. The new value will be used internally
immediately (even before cycling power off and back on), but the
IPNAME? query will return the previous system name until the
Model 430 Programmer is restarted.
•
IPNAME?
Returns the system name (also known as
host name
or
computer
name
).
R
AMP
T
ARGET
/R
ATE
C
ONFIGURATION
C
OMMANDS
AND
Q
UERIES
The ramp configuration commands set the various parameters required
for defining and limiting piecewise-linear ramp segments.
•
CONFigure:VOLTage:LIMit <
voltage (V)
>
Sets the ramping Voltage Limit in volts. The limit may not exceed the
maximum output voltage of the power supply.
Summary of Contents for 430
Page 2: ......
Page 16: ...XVI REV 9 AMERICAN MAGNETICS INC FOREWORD SAFETY SUMMARY ...
Page 28: ...12 REV 9 AMERICAN MAGNETICS INC INTRODUCTION OPERATING CHARACTERISTICS ...
Page 64: ...48 REV 9 AMERICAN MAGNETICS INC INSTALLATION POWER UP AND TEST PROCEDURE ...
Page 208: ...192 REV 9 AMERICAN MAGNETICS INC SERVICE RETURN AUTHORIZATION ...
Page 248: ...232 REV 9 AMERICAN MAGNETICS INC APPENDIX SHORT SAMPLE MODE ...