1. Using serial null modem adapter (and possibly a gender changer)
5
, connect
the DB9 RS-232 connector on the rear of the Model 430 Programmer to a
standard USB-to-serial cable
6
connected to a computer.
2. Start a terminal emu-
lator program on the
remote computer. As
an example, this pro-
cedure will use the
open-sourced Tera
Term program
7
run-
ning on a Windows
machine. You are
greeted immediately
with the
Tera Term
New Connection
dia-
log.
3. Choose the
Serial
option and then the
Port
to which the Model 430 is con-
nected. This example illustrates
COM20: USB Serial Port (COM20),
which in
this example is the port assigned by Windows to a USB-to-serial adapter
cable. Press
OK
.
4. Use the
Setup | Serial
Port...
menu com-
mand in Tera Term to
show the serial
parameters dialog.
Set the parameters
as shown at right per
page 143 (
Port
selec-
tion is specific to your
setup).
5. Mouser Electronics offers a compact null-model/gender changer: see part # 515-140-448-R.
6. AMI recommends FTDI USB-to-serial cables. See DigiKey part # 768-1084-ND.
7. The Tera Term application is open-sourced and is available at https://ttssh2.osdn.jp/
index.html.en. A Windows installer for a
recent
version (not guaranteed to be the latest) is
also included on the internal Model 430 web page, see page 226.
Summary of Contents for 430
Page 2: ......
Page 16: ...XVI REV 9 AMERICAN MAGNETICS INC FOREWORD SAFETY SUMMARY ...
Page 28: ...12 REV 9 AMERICAN MAGNETICS INC INTRODUCTION OPERATING CHARACTERISTICS ...
Page 64: ...48 REV 9 AMERICAN MAGNETICS INC INSTALLATION POWER UP AND TEST PROCEDURE ...
Page 208: ...192 REV 9 AMERICAN MAGNETICS INC SERVICE RETURN AUTHORIZATION ...
Page 248: ...232 REV 9 AMERICAN MAGNETICS INC APPENDIX SHORT SAMPLE MODE ...