Operating Manual GC 5000 BTX; Version 5 – 11/2010
77
11.3.4 Master mode
Menu bar
Parameter
Master mode
If you have an ozone precursors system (one GC 5000 BTX analyzer and
one GC 5000 VOC analyzer) you can switch on the master mode and set
the VOC analyzer into slave mode.
In this case the BTX analyzer is the master and controls the VOC analyzer
(the slave). Each cycle and calibrations are started by the BTX analyzer. So
the two analyzers are running synchronously.
Note
It is necessary that both analyzers have the same operation mode, the
same number of calibration levels and the same calibration setup. In case a
complete ozone precursor system (GC 5000 BTX and GC 5000 VOC) is
manufactured at the same time by AMA Instruments, the parameters are
configured ex factory .
11.3.5 Method
Menu bar
Parameter
Method
The current set of parameters can be saved as a method or a new set of
parameters can be loaded.
The following window pops up:
1. Select a saved method and after clicking onto “Open” it is loaded.
2. Analog you can save a set of parameters as a method with clicking onto
“Save”.