Operating Manual GC 5000 BTX; Version 5 – 11/2010
145
Field no.
Start byte
Data format
Field description
19
41
b
Serial number
20
42
b
Monitoring instrument ID
instrument 4 (optional)
21
43
±
nnnn
±
ee
Measured value
22
51
b
Operating status
23
52
b
Error status
24
53
b
Serial number
25
54
b
Monitoring instrument ID
instrument 5 (optional)
26
55
±
nnnn
±
ee
Measured value
27
63
b
Operating status
28
64
b
Error status
29
65
b
Serial number
30
66
b
Monitoring instrument ID
instrument 6 (optional)
31
67
±
nnnn
±
ee
Measured value
32
75
b
Operating status
33
76
b
Error status
34
77
b
Serial number
35
78
b
Monitoring instrument ID
instrument 7 (optional)
36
79
±
nnnn
±
ee
Measured value
37
87
b
Operating status
38
88
b
Error status
39
89
b
Serial number
40
90
b
Monitoring instrument ID
instrument 8 (optional)
41
91
±
nnnn
±
ee
Measured value
42
99
b
Operating status
43
100
b
Error status
44
101
b
Serial number
45
102
<ETX>
End of text
46
103
<BCC1>
Upper nibble
47
104
<BCC2>
Lower nibble
From 21 or more monitoring instruments there are two length bytes:
Field no.
Start byte
Data format
Field description
1
1
<STX>
Start of text
2
2
bb
Length bytes (Upper byte, lower
byte)
3
4
AD
Telegram identification
4
6
b
Number of monitoring
instruments to transmit data
5
7
b
Monitoring instrument ID
instrument 1
6
8
±
nnnn
±
ee
Measured value
7
16
b
Operating status
8
17
b
Error status
9
18
b
Serial number
…
Summary of Contents for The GC 5000 BTX
Page 1: ...Operating Manual Gas chromatograph GC 5000 BTX...
Page 35: ...Operating Manual GC 5000 BTX Version 5 11 2010 33 Fig 5 Connection of analog outputs...
Page 102: ...Operating Manual GC 5000 BTX Version 5 11 2010 100 Injection Waiting for end of cycle time...
Page 105: ...Operating Manual GC 5000 BTX Version 5 11 2010 103 Injection Waiting for end of cycle time...