Operating Manual GC 5000 BTX; Version 5 – 11/2010
26
Note!
After switching on the analyzer, it is possible that the enrichment stage
heats up to 230°C. This is not a malfunction. It cools down immediately to
the specified enrichment temperature.
Note!
When using synthetic air with a very high purity instead of air from a zero
air generator as combustion air make sure that the FID signal-offset of the
FID amplifier is adjusted correctly (see chapter 13.3.5). The FID signal may
be lower with synthetic air than with zero air. If the signal is still < 100 mV
when the flame is lit, you will get an alarm message (see chapter 14.2 –
alarm 5.3).
Note
The colored field indicates the status of the system.
Green
: the analyzer is
ready to start and runs problem-free (see fig. below).
Yellow
: the system is
not ready as set points of the temperatures have not been reached.
Red
:
An alarm or error has occurred.