Operating Manual GC 5000 BTX; Version 5 – 11/2010
140
Type of Error
Cause
Information
Error 11:
“Communication
breakdown to sample
MFC. The program has
been stopped!”
Built-in computer gets no
connection to the sample
MFC.
This means that current sample flow values are not
shown on the display anymore.
The flow set points cannot be changed.
Error 13:
“Communication
breakdown to
hydrogen MFC. The
program has been
stopped!”
Built-in computer gets no
connection to the
hydrogen MFC.
This means that current hydrogen flow values are not
shown on the display anymore.
Flow set points cannot be changed.
15 Spare parts list
The list below contains the main components. They are necessary
for maintenance which will be performed on a preventive basis.
Component
Art.-No
Back panel fan
1253
BTX enrichment tube
220
Fan of the Peltier cooler
131
GC column
1310
GC oven motor
1254
Inlet filter for MFC
427
Maintenance kit FID (FID jet, sealings)
10000
Maintenance kit PID (lamp, centering ring, sealing)
2013
Pump
526
Rotor of the rotary valve
242