Operating Manual GC 5000 BTX; Version 5 – 11/2010
62
3.
Check validation status:
Select Menu bar
View
Validation status
In this window the result is displayed after a validation sequence has
completed. In the example below every check is within the defined limits.
•
Green indicates that the validation run of all tests is ok.
•
Yellow indicates that the limit is reached (warning level)
•
Red indicates that the values exceed the limit (error level). For further
information and definition see page 60.
3. Check the exact results by opening e.g. the file
c:\cpdata\gc5000\VALBenzene.btx.met.txt
in case Benzene has been used for validations.
Note
The name of the file consists of VAL + component name + method name +
.txt.
The columns of the file contain the following information:
•
Counter
•
“I” for initial validation / “S” for standard validation
•
Name of chromatogram for zero gas check
•
Date and time of chromatogram for zero gas check
•
Result of chromatogram for zero gas check
•
Name of chromatogram for drift check
•
Date and time of chromatogram for drift check
•
Result of chromatogram for drift check
•
Name of chromatogram for deviation check
•
Date and time of chromatogram for deviation check
•
Result time of chromatogram for deviation check
•
Drift
•
Deviation
Structure of
file:
Summary of Contents for The GC 5000 BTX
Page 1: ...Operating Manual Gas chromatograph GC 5000 BTX...
Page 35: ...Operating Manual GC 5000 BTX Version 5 11 2010 33 Fig 5 Connection of analog outputs...
Page 102: ...Operating Manual GC 5000 BTX Version 5 11 2010 100 Injection Waiting for end of cycle time...
Page 105: ...Operating Manual GC 5000 BTX Version 5 11 2010 103 Injection Waiting for end of cycle time...