Altec AIR, LLC
P210WLP Series Air Dryer User’s Guide
Page 23 of 96
P017455 – Rev. N
6.5.14
Set the System Pressure:
With Compressor running:
6.5.14.1
Pull the Capacity Control Valve
knob out.
6.5.14.2
Turn the knob until the reading
on the pressure gauge is
50 PSI
.
6.5.14.3
Push the knob in to lock.
6.5.15
Let the Dryer run until the Humidity drops below 2%. (may take up to 15
minutes).
NOTE
: Press the
RESET
button if the Dryer goes into
SHUTDOWN
mode.
6.5.16
Power the Dryer
OFF
.
6.5.17
Remove the Precision Bleed Orifice fitting from the Outlet Port by
pressing the ferrule down then pulling the fitting upward.