97
-EN
Calibration when connected to
HCE-C300R
Obstacle detection performance adjustment and guidance mark settings
of HCE-C300R (REAR VIEW CAMERA SYSTEM) require that
camera installation information be entered and calibration performed. If
camera information is not entered, accurate detection will not be
possible, so be sure to perform calibration.
For details on calibration, refer to “Calibration” in the owner’s manual
of HCE-C300R. To perform calibration with the supplied switch, refer
to “Performing Calibration” in the owner’s manual of HCE-C300R.
When connected to this unit, you can perform operations via the screen
of this unit.
To activate the Calibration Setup Mode in this unit
1
Perform steps 1 and 2 in “Performing Calibration” of
the owner’s manual of HCE-C300R.
2
Press
SOURCE/
on the unit.
The SOURCE selection screen is displayed.
3
Touch
(CAMERA).
The Camera mode is activated and the Top menu is
displayed.
4
Touch any area of the screen to display the view
change function keys.
•
Make sure view change function is set to ON (page 48).
Display example for Top menu screen
5
Touch
[Back]
[Corner]
[Back]
according to
the instructions on the screen.
The Setup Mode screen appears.
Display example for Setup mode screen
Continue with the following operations after performing step 4 in
“Performing Calibration” of the owner’s manual of HCE-C300R. While
doing such operations, touch the button on the screen to select
“CHANGE” or “DONE” to perform calibration on this unit. Touching
[Corner]
and
[Back]
on the screen of this unit performs the same
operation as pressing the switch which is supplied by HCE-C300R.
[Corner] = VIEW:
Selects a desired item.
(The switch that is selected is enclosed by a
yellow frame.)
[Back] = SET.:
Executes the selection.