![Alcatel Vacuum Technology Adixen ATP 100 User Manual Download Page 65](http://html.mh-extra.com/html/alcatel-vacuum-technology/adixen-atp-100/adixen-atp-100_user-manual_2891737065.webp)
8/10
B 120
Alcatel Vacuum Technology France - ATP User’s Manual
Detailed description of RS commands
GB 00420 - Edition 03 - March 00
Syntax
Result
SEL
Syntax
Result
Returns the state of the parameters defined by OPT
#adr
SEL
<CR>[<LF>]
#adr,a,u
a : Returns choice of parameters on the analog output:
a = 0 : real pump speed
a = 1 : pump current
a = 2 : temperature of pump body
a = 3 : temperature of internal electronics
u : Returns the choice of temperature unit:
u = 0 : degrees Centigrade
u = 1 : degrees Fahrenheit
Complete cabinet only
#adr
SEL
10<CR>[<LF>]
#adr,a,u,s,b
a : Returns choice of parameters on the analog output:
a = 0 : real pump speed
a = 1 : pump current
a = 2 : temperature of pump body
a = 3 : temperature of internal electronics
u : Returns the choice of temperature unit:
u = 0 : degrees Centigrade
u = 1 : degrees Fahrenheit
s : Returns auto-starting choice:
s = 0 : no
s = 1 : yes
b : Returns buzzer choice:
b = 0 : without
b = 1 : with
Defines the character which separates the parameters
in a reply
#adr
SEP
nnn<CR>[<LF>]
nnn: 3-digit decimal value of the ascii code of the desired
character (with leading zeros).
condition : 000
≤
nnn
≤
255
#adr,ok
or
#adr,ErrX if error
Allows the user to select the character which separates
the parameters returned by the
DLR
,
STA
and
LEV
commands. Default value: comma «,» ascii code = 044
If ok, the selected value is automatically stored in user memory.
SEP
Syntax
Result