![Alcatel Vacuum Technology Adixen ATP 100 User Manual Download Page 58](http://html.mh-extra.com/html/alcatel-vacuum-technology/adixen-atp-100/adixen-atp-100_user-manual_2891737058.webp)
1/10
B 120
Alcatel Vacuum Technology France - ATP User’s Manual
GB 00420 - Edition 03 - March 00
Detailed description of RS commands
(valid from V1.10 version variable drives)
Conventions applicable
to the syntax of all
commands:
Status values
Error messages
ADR
Syntax
Result
BRK
Syntax
Result
adr
= address, from 000 to 255
<CR>
Carriage Return (ascii 13)
<LF>
Line Feed (ascii 10); between square brackets:
this character is not compulsory.
ok
: command executed correctly
Err0
: adjustment error (out of bounds)
Err1
: command error (syntax)
Err2
: parameter error (e.g. non-hexadecimal character)
Err3
: context error
Err4
: checksum error
Specifies the address of the device for networking.
#adr
ADR
aaa<CR>[<LF>]
adr
= address of the device before the command
aaa
= new address of the device
condition
: 000
≤
aaa
≤
255
#aaa,ok or Err2
This command is used to allocate a specific number to
each of the products making up a network
(loop for RS 232 or parallel for RS 485).
Note : it is important to note down the number
allocated to each device.
Stop the pump by braking
(ATP 80/100 series only)
#adr
BRK
<CR>[<LF>]
#adr,ok
This command is used to brake the motor electrically, which is
particulary effective at high speed. It is currently only available for