Alcatel Vacuum Technology France - ATP User’s Manual
GB 00439 - Edition 01 - March 97
ATP 150, ATP 400 and ATP 900
Pumps lubrication
2/2
Unfasten the lubricating
plug on the end cap side.
Introduce the syringe equipped
with its needle to the bottom
of the housing and
remove
the red jumper.
While keeping the syringe
needle at the bottom of its
housing, introduce the grease
until the plunger comes to a stop. Remove the syringe.
Replace the lubricating plug after replacing its o-ring,
contained in the syringe packaging.
Unfasten the lubricating
plug on the pumping cell side.
Repeat the same operations,
this time removing the black
jumper from the syringe (keep
the syringe at the bottom of
its housing throughout the
operation).
Remove the syringe and reassemble the lubricating plug
equipped with its new o-ring.
The relubrication operation is complete.
Execute the pump running-in:
see
E 70
for ATP 150 and ATP 400, and
see
E 90
for ATP 900.
13
13
Lubrication of the bearing
opposite the pumping cell
Lubrication of the bearing
on the pump cell side
End cap
The pump can remain connected to the installation during
lubrication.