Alcatel Vacuum Technology France - ATP User’s Manual
Detailed description of RS commands
GB 00420 - Edition 03 - March 00
RPM
Syntax
Result
See also:
NSP, SBY
Defines the speed set point in stand-by mode
#adr
RPM
nnnnn<CR>[<LF>]
or
#adr
RPM
nnnnn<CR>[<LF>]
#adr,ok
or
#adr,ErrX
1, out of range; 2, parameters ; 3, context (not in Stand-by mode)
Comment: if ok, the new speed is automatically stored in user memory.
Saves the internal parameters in user’s memory
#adr
SAV
<CR>[<LF>]
#adr,ok
Saves the current context (except for running-in cycles).
If this command is sent when the pump is being supplied,
it can for example allow automatic re-start in the event of
a power cut.
Switches the speed set point to the stand-by value
#adr
SBY
<CR>[<LF>]
#adr,ok
Resets the stand-by speed to its last stored value, and allows
it to be modified if an «RPM» command is sent.
This configuration is automatically stored in user memory.
SAV
Syntax
Result
SBY
Syntax
Result
See also:
NSP, RPM