![Alcatel Vacuum Technology Adixen ATP 100 User Manual Download Page 45](http://html.mh-extra.com/html/alcatel-vacuum-technology/adixen-atp-100/adixen-atp-100_user-manual_2891737045.webp)
B 60
ACT 200 T and 600 T
"Remote Control" connector wiring
Alcatel Vacuum Technology France - ATP User’s Manual
2/2
GB 00414 - Edition 03 - February 99
Signaling using
output contacts:
These are dry contacts (
250VAC-1A),
their function is to
replicate the data concerning the pump operating status.
When the output
contact
is closed, this indicates:
15
14
13
12
11
10
9
8
7
1
30
16
44
43
31
42 41
40
39
OV(-)
OV
+ 15 V Max 50 mA
10V(+)
44 - 43
Used to monitor the selected parameter in
the “Set Analog output“ menu (
see B 70
).
A 0-10V analog output
is used to monitor variations in
certain pump parameters (speed, temperature, etc.)
This data can be used to plot curves.
1 - 2
The pimp is in running-in mode.
5 - 6
The pump is operating
7 - 8
Copy of Start command.
This contact can be used to drive
the roughing pump (
see C 50
).
9 - 10
The standby mode is selected.
11 - 12
No faults are signalled.
13 - 14
The pump has reached its nominal speed.
15 - 30
The venting valve fitted in the pump is closed
(accessory).