B 50
ACT 200 T and 600 T
electrical connections
Alcatel Vacuum Technology France - ATP User’s Manual
1/1
GB 00413 - Edition 02 - February 99
0
1
RS 232 / 485
TURBO PUMP CABLE
REMOTE CONTROL
If the controller is remote controlled, make the various
connections on the
REMOTE CONTROL
connector
(see
B 60
).
If the controller unit is
used in
local mode:
the pump can only
operate
if the plug is connected
to the REMOTE CONTROL
connector
.
(Connector plug delivered with
the controller).
Connect the
RS232/485
serial link cable to the
connector
(cable supplied by customer)
Fuse access
15
1
30
16
44
41
39
31
REMOTE CONTROL
connector plug
(factory wiring)
Wiring viewed on soldered side
Connect
the controller to the
pump
using
the cable
ordered.
Connect
the controller to
the mains using the cable
supplied.
(Mains with earth
connection)
Connect the fan
(ATP air cooled)
Connect the
electrical venting valve
powered by the mains and
driven by the controller.
(voir
B 60
).
Power
supply
Delayed fuses
ACT 200T
3.15 A
250 V
ACT 600T
6.3 A
250 V
Display contrast setting