Application Note CC-002-A
90
Figure 3B. Connections for Dual Posts.
TRIPLE INTERCONNECTIONS
Figures 4 and 5 show the typical intercell connections for triple post
cells. For cells arranged as in Fig. 4, three readings are made. The first
one, shown, should be with the intercell leads connected from Terminal
Post A to Terminal Post D. Take the second reading with the intercell
leads connected from Terminal Post B to Post E. The third reading is
with the intercell leads connected from Terminal Post C to Post F.