CRT-300 Specifications
73
Environmental
Temperature Range
Operating
32
F to 104
F (0
C to 40
C)
Storage
–20
F to 140
F (–29
C to 60
C)
For extended periods: <77
F (<25
C)
Humidity Range
0% to 80%RH (non-condensing) at 0
C to 31
C
0% to 50%RH (non-condensing) at 32
C to 40
C
Insulation Category II
Pollution Degree 2
Altitude
0 to 2000 meters above sea level
Accuracy
Input
Test
Current
Resistance
Resolution
0.1% of reading
Voltage
Resolution
0.1% of reading
Intercell
Range
Resolution
1% of reading
2V
60A
5
1mV
2m
4V to 6V
25A to 35A
25
3mV
3.3m
8V to 12V
15A to 30A
75
6mV
5m
16V
5A
400
10mV
20m
Warranty
One year limited warranty
Ordering Information
CRT-300
Cellcorder, including software, carrying case, battery
charger, cables and clamps
Optional Accessories
2025-124
Printer with IRDA interface
2025-031
Internal modem
2030-002
Smart Media card, 128MB
Note: Specifications are subject to change without notice.