13 Information on scattered
radiation
13.1 CBCT scattered radiation
Test equipment: Dosimeter Radcal 9015
Test conditions
Program parameters
CBCT
Image acquisition vol-
ume
Normal
Voltage
99 kVp
Current
14 mA
R
1 m
1.5 m
2 m
°
mR/h
mR/h
mR/h
0
588.2
135.3
87.1
45
549.3
249.4
106.8
90
472.6
307.3
78.4
135
458.8
287.6
89.3
180
12.9
4.6
1.3
225
410.5
288.7
98.2
270
663.2
301.4
112.4
315
429.7
194.2
92.3
13.2 Pano scattered radiation
Test equipment: Dosimeter Radcal 9015
Test conditions
Program parameters
Panoramic Standard
Patient size
Normal
Voltage
80 kVp
Current
14 mA
R
1 m
1.5 m
2 m
°
mR/h
mR/h
mR/h
0
60.9
17.7
8
45
19.6
12.4
5.8
90
10.6
6.8
4.1
135
22.1
12.5
5.6
180
1
0
0
225
45.4
21.4
9.4
270
47.6
21.9
9.2
R
1 m
1.5 m
2 m
°
mR/h
mR/h
mR/h
315
76.4
19.4
8.6
Appendix
52
2210200496L41 1911V006