After the acquisition of maxillary joint
images, it is necessary to confirm a
message on the touch screen and trig-
ger a second image acquisition. The
images are then combined into a single
image.
The LED on the unit lights up blue when the X-
ray acquisition has been completed.
The rotating unit does not automatically move
back to the starting position after the trigger
button is released.
❯
Click
OK
to confirm the message.
❯
Release the head supports.
The patient can leave the X-ray room.
❯
Remove the hygienic protective cover.
❯
Remove and disinfect the positioning aids.
❯
The unit can be returned to its starting position
by touching
Start position
. Otherwise, the C-
shaped arm is positioned via the imaging soft-
ware during adjustment of the parameters.
7.7
Emergency stop switch
The emergency stop switch stops the unit and
switches it off. It can be used if the unit is taking
an X-ray even though the exposure button is no
longer being pressed, or if the patient is injured or
the unit is damaged. It can also be used to avert
an unwanted collision.
The yellow stickers on the patient positioning
system with the symbol
show the location of
the emergency stop switch.
Fig. 4: Emergency stop switch sticker on the
operator side
Fig. 5: Emergency stop switch sticker on the
patient side
Misuse of the emergency stop switch can
lead to data loss.
Usage
38
2210200496L41 1911V006
EN-
US