5.3
Status LED
The status LED uses different colors to display
the different operating modes:
– Blue: unit ready for operation
– Green: unit ready to acquire image
– Yellow: X-ray beam active
5.4
Positioning aids
3
1 2
5
7
6
4
8
1
Lever for positioning the Frankfurt plane
positioning beam
2
Frankfurt plane of the X-ray positioning
beam
3
Head supports with cushion
4
Positioning aid, e. g. chin support with bite
block
5
Upper canine positioning beam
6
Lever for positioning the upper canine posi-
tioning beam
7
Mid-sagittal positioning beam
8
Grips
The applied parts in accordance with IEC
60601-1 are:
– Grips
– Head supports with cushion
– Positioning aids (e.g. bite block and mounting
for bite block, chin support for edentulous
patients)
Description of the positioning aids
The positioning aids are used to correctly posi-
tion the patient in the unit. The appropriate posi-
tioning aid is selected according to the selected
image. The head supports gently keep the head
of the patient in place.
Panoramic
Bite block and adapter bite
block
Chin holder for edentulous
jaws
Chin holder for maxillary joint
image
Chin holder for sinus image
Head supports with cushion
Product description
2210200496L41 1911V006
19
EN-
US
Summary of Contents for Rxonly ProVecta 3D Prime
Page 1: ...EN US ProVecta 3D Prime Operating instructions 2210200496L41 2210200496L41 A7766 1911V006...
Page 2: ......
Page 26: ...Narrow jaw arch Wide jaw arch Child s jaw arch Usage 24 2210200496L41 1911V006 EN US...
Page 56: ......
Page 57: ......
Page 58: ......
Page 59: ......