Perform the following steps:
❯
Sterilize the parts for sterilization (at least 20
minutes at 250°F, at least 4 minutes at 270°F
or at least 5 minutes at 274°F).
Do not exceed 281 °F.
Marking
❯
Mark the packaged, treated medical device
appropriately such as to ensure its safe appli-
cation.
9.9
Issue clearance for the parts
for sterilization
The reprocessing of the medical device ends with
the documented clearance for storage and
renewed use.
❯
Document the release of the medical device
after reprocessing.
9.10
Storing parts for sterilization
❯
Comply with the stated storage conditions:
– Store the parts protected against contami-
nation
– Dust-protected, e.g. in a locked cabinet
– Protected against moisture
– Protected against excessive temperature
fluctuations
– Protected against damage
The integrity of the packaging of a sterile medi-
cal device be lost as a result of a particular inci-
dent and the passage of time.
Potential external contamination of the sterile
barrier system should be taken into account in
terms of aseptic preparation when establishing
the storage conditions.