Appendix
12 Program parameters
The digital extraoral dental X-ray system meets the requirements set out in standard IEC 60601-2-63.
The dosage information complies with the requirements of the standard and is stated in mGy.
The accuracy of the DAP/dose values is
±
50%.
12.1 CBCT program parameters
CBCT image acquisition, normal image volume, 16.4 s
4.0 mA
6.3 mA
8.0 mA
10.0 mA
mGy
mGycm
2
mGy
mGycm
2
mGy
mGycm
2
mGy
mGycm
2
75 kV
4.00
215.92
6.13
331.05
7.79
420.70
9.74
526.41
79 kV
4.50
242.90
6.89
372.41
8.76
473.26
10.96
592.17
90 kV
6.00
324.11
9.20
496.93
11.69
631.50
14.63
790.18
94 kV
6.55
353.92
10.04
542.63
12.77
689.58
15.97
862.85
CBCT image acquisition, normal image volume 5x5, 11 s
4.0 mA
6.3 mA
8.0 mA
10.0 mA
11.0 mA
mGy
mGyc
m
2
mGy
mGyc
m
2
mGy
mGyc
m
2
mGy
mGyc
m
2
mGy
mGyc
m
2
79 kV
3.02
118.85 4.63
182.22 5.88
231.56 7.36
289.75 8.08
318.22
94 kV
4.40
173.17 6.74
265.51 8.57
337.41 10.72
422.19 11.78
463.68
98 kV
4.77
187.76 7.31
287.87 9.29
365.83 11.63
457.75 12.77
502.73
12.2 Panoramic program parameters
Pano image acquisition, normal jaw arch, normal patient, quality HQ, 13.5 s
4.0 mA
6.3 mA
10.0 mA
mGy
mGycm
2
mGy
mGycm
2
mGy
mGycm
2
60 kV
3.85
24.28
6.06
38.15
9.57
60.31
67 kV
4.74
29.84
7.43
46.83
11.76
74.09
70 kV
5.12
32.24
8.03
50.59
12.70
80.03
74 kV
5.66
35.66
8.88
55.95
14.05
88.52
80 kV
6.47
40.79
10.16
64.00
16.07
101.25
12.5 mA
14.0 mA
mGy
mGycm
2
mGy
mGycm
2
60 kV
11.77
74.18
13.21
83.23
67 kV
14.46
91.07
16.22
102.21
70 kV
15.61
98.37
17.52
110.40
Appendix
50
2210200496L41 1911V006
EN-
US
Summary of Contents for Rxonly ProVecta 3D Prime
Page 1: ...EN US ProVecta 3D Prime Operating instructions 2210200496L41 2210200496L41 A7766 1911V006...
Page 2: ......
Page 26: ...Narrow jaw arch Wide jaw arch Child s jaw arch Usage 24 2210200496L41 1911V006 EN US...
Page 56: ......
Page 57: ......
Page 58: ......
Page 59: ......