Chapter 4 Remote Interface Reference
The SCPI Status Registers
95
4
The Questionable Status Register
The Questionable Status register provides information about voltage and
current regulation. Bit 0 is set when the voltage becomes unregulated, and bit
1 is set if the current becomes unregulated. For example if the power supply
momentarily goes to constant current mode when the power supply is
operating as a voltage source (constant voltage mode), bit 0 is set to indicate
that the voltage output is not regulated.
The Questionable Status register also provides information that the power
supply has an overtemperature condition and that the overvoltage protection
circuits have tripped. Bit 4 reports an overtemperature condition of the fan and
bit 9 reports that the overvoltage protection circuit has tripped. To read the
register, send
STATus:QUEStionable?
.
Table 4-3. Bit Definitions - Questionable Status Register
The Questionable Status Event register is cleared when:
• You execute the
*CLS
(clear status) command.
• You query the event register using
STAT:QUES?
(Status Questionable
Event register) command.
For example, 16 is returned when you have queried the status of the
questionable event register, the temperature condition is questionable.
The Questionable Status Enable register is cleared when:
• You execute
STAT:QUES:ENAB 0
command.
Bit
Decimal
Value
Definition
0
Voltage
1
The power supply is/was in the constant current mode.
1
Current
2
The power supply is/was in the constant voltage mode.
2-3
Not Used
0
Always set to 0.
4
Overtemperature
16
The fan has a fault condition.
5-8
Not Used
0
Always set to 0.
9
Over Voltage
512
The overvoltage protection circuit has tripped.
10
Not Used
0
Always set to 0.
11-15 Not Used
0
Always set to 0.
Summary of Contents for E364 A Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 27: ...Chapter 1 Quick Start To Rack Mount the Instrument 26...
Page 28: ...2 General Information...
Page 40: ...3 Front Panel Operation and Features...
Page 68: ...4 Remote Interface Reference...
Page 114: ...5 Error Messages...
Page 124: ...6 Application Programs...
Page 135: ...Chapter 6 Application Programs Example Program for Excel 97 134...
Page 136: ...7 Tutorial...
Page 147: ...Chapter 7 Tutorial Remote Programming 146...
Page 148: ...8 Specifications...
Page 156: ...Appendix Service Information...
Page 165: ...Appendix Service Information General Disassembly 164 General Disassembly...
Page 193: ...Appendix Service Information Replaceable Parts 192...
Page 197: ......
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......