Appendix Service Information
Calibration Record for Your Power Supply
187
Ser
vice
In
for
mat
ion
Calibration Record for Your Power Supply
Step
Calibration Description
Measurement
Mode (DVM)
Supply Being
Adjusted
1
Turn on the calibration mode by holding down the "Calibrate" key as you
turn on the power supply until you hear a long beep.
2
Unsecure the power supply if secured. (See page 62)
3
Press "Calibrate" key to move down menu to voltage calibration menu.
A "
VOLTAGE CAL
" is displayed. Press "Calibrate" key to select the
low voltage calibration point.
Voltage
Calibration
4
A low voltage calibration point is displayed. Enter the DVM reading by
using the knob and resolution keys. Press "Calibrate" key to save the
changes and select the middle calibration point.
V
Low voltage
calibration
5
A middle voltage calibration point is displayed. Enter the DVM reading
by using the knob and resolution selection keys. Press "Calibrate" key to
save the changes and select the high calibration point.
V
Middle voltage
calibration
6
A high voltage calibration point is displayed. Enter the DVM reading by
using the knob and resolution selection keys. Press "Calibrate" key to
save the changes and select the OVP calibration.
V
High voltage
calibration
7
"OVP CAL" is displayed. Press "Calibrate" key to run the overvoltage
calibration. A "CALIBRATING" is displayed to indicate that the
calibration is progressing.
V
OVP calibration
8
Connect shunt resistor (0.01
9
) across the output terminals. And press
"Calibrate" key to select the current calibration.
Current calibration
9
A low current calibration point is displayed. Enter the computed value
(DVM reading / by shunt resistance) by using the knob and resolution
keys. Press "Calibrate" key to save the changes and select the middle
calibration point.
A
Low current
calibration
10
A middle current calibration point is displayed. Enter the computed value
(DVM reading / by shunt resistance) by using the knob and resolution
keys. Press "Calibrate" key to save the changes and select the high
calibration point.
A
Middle current
calibration
11
A high current calibration point is displayed. Enter the computed value
(DVM reading / by shunt resistance) by using the knob and resolution
keys. Press "Calibrate" key to save the changes
A
High current
calibration
13
Turn off the "Power" switch to exit the calibration menu.
Exit CAL MODE
Summary of Contents for E364 A Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 27: ...Chapter 1 Quick Start To Rack Mount the Instrument 26...
Page 28: ...2 General Information...
Page 40: ...3 Front Panel Operation and Features...
Page 68: ...4 Remote Interface Reference...
Page 114: ...5 Error Messages...
Page 124: ...6 Application Programs...
Page 135: ...Chapter 6 Application Programs Example Program for Excel 97 134...
Page 136: ...7 Tutorial...
Page 147: ...Chapter 7 Tutorial Remote Programming 146...
Page 148: ...8 Specifications...
Page 156: ...Appendix Service Information...
Page 165: ...Appendix Service Information General Disassembly 164 General Disassembly...
Page 193: ...Appendix Service Information Replaceable Parts 192...
Page 197: ......
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......