Chapter 1 Quick Start
Output Checkout
19
1
Current Output Checkout
The following steps check basic current functions with a short across the
power supply’s output.
1 Turn on the power supply.
Make sure that the output is disabled. The
OFF
annunciator is on
2 Connect a short across (+) and (-) output terminals with an insulated
test lead.
Use a wire size sufficient to handle the maximum current (See "
Table 2-1 Wire
Rating
" on page 34 in chapter 2).
3 Enable the output.
The
CV
or
CC
annunciator turns on depending on the resistance of the test
lead. Notice that the display is in the
meter
mode.
4 Adjust the voltage limit value to 1.0 volt.
Set the display to the
limit
mode (the
Limit
annunciator will be flashing).
Adjust the voltage limit to 1.0 volt to assure CC operation. The
CC
annunciator
will turn on. To go back to normal mode, press the
key again or let the
display time out after several seconds.
5 Set the knob to the current control to check that the front-panel
ammeter properly responds to knob control.
Turn the knob clockwise or counter clockwise when the display is in the
meter
mode (the
Limit
annunciator is off). Check that the ammeter responds to knob
control and the voltmeter indicates nearly zero (the voltmeter will show the
voltage drop caused by the test lead). The
flashing
digit can be adjusted by
turning the knob.
6 Ensure that the current can be adjusted from zero to the full rated
value.
7 Turn off the power supply and remove the short from the output
terminals.
1
You can use the resolution selection keys to move the flashing digit to the
right or left when setting the current.
Power
On/Off
Output
Display
Limit
Display
Limit
Current
Voltage
1
Summary of Contents for E364 A Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 27: ...Chapter 1 Quick Start To Rack Mount the Instrument 26...
Page 28: ...2 General Information...
Page 40: ...3 Front Panel Operation and Features...
Page 68: ...4 Remote Interface Reference...
Page 114: ...5 Error Messages...
Page 124: ...6 Application Programs...
Page 135: ...Chapter 6 Application Programs Example Program for Excel 97 134...
Page 136: ...7 Tutorial...
Page 147: ...Chapter 7 Tutorial Remote Programming 146...
Page 148: ...8 Specifications...
Page 156: ...Appendix Service Information...
Page 165: ...Appendix Service Information General Disassembly 164 General Disassembly...
Page 193: ...Appendix Service Information Replaceable Parts 192...
Page 197: ......
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......