Chapter 3 Front-Panel Operation and Features
Programming Overvoltage Protection
52
1 Raise the OVP trip level higher than the level tripped.
2 Select the OVP CLEAR mode by turning the knob.
3 Clear the overvoltage condition and exit this menu.
The
OVP
annunciator will not flash any more. The output will return to the
meter mode.
•
Remote interface operation:
VOLT:PROT {<voltage>|MIN|MAX}
Set the OVP level
VOLT:PROT:STAT {OFF|ON)
Disable or enable the OVP circuit
VOLT:PROT:CLE
Clear the tripped OVP circuit
By Adjusting OVP trip level
OVP
OVP
OVP
OVP
ON
ON
ON
ON
OVP
OVP
OVP
OVP
CLEAR
done
done
done
done
Over
Voltage
Over
Voltage
Over
Voltage
Summary of Contents for E364 A Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 27: ...Chapter 1 Quick Start To Rack Mount the Instrument 26...
Page 28: ...2 General Information...
Page 40: ...3 Front Panel Operation and Features...
Page 68: ...4 Remote Interface Reference...
Page 114: ...5 Error Messages...
Page 124: ...6 Application Programs...
Page 135: ...Chapter 6 Application Programs Example Program for Excel 97 134...
Page 136: ...7 Tutorial...
Page 147: ...Chapter 7 Tutorial Remote Programming 146...
Page 148: ...8 Specifications...
Page 156: ...Appendix Service Information...
Page 165: ...Appendix Service Information General Disassembly 164 General Disassembly...
Page 193: ...Appendix Service Information Replaceable Parts 192...
Page 197: ......
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......