Chapter 4 Remote Interface Reference
The SCPI Status Registers
96
The Standard Event Register
The Standard Event register reports the following types of instrument events:
power-on detected, command syntax errors, command execution errors, self-
test or calibration errors, query errors, or when an
*OPC
command is executed.
Any or all of these conditions can be reported in the standard event summary
bit (ESB, bit 5) of Status Byte register through the enable register. To set the
enable register mask, you write a decimal value to the register using the
*ESE
(Event Status Enable) command.
Table 4-4. Bit Definitions – Standard Event Register
An error condition (Standard Event register bit 2, 3, 4, or 5) will always
record one or more errors in the power supply’s error queue. Read the
error queue using the
SYST:ERR?
command.
Bit
Decimal
Value
Definition
0
OPC
1
Operation Complete. All commands prior to and
including an
*OPC
command have been executed.
1
Not Used
0
Always set to 0.
2
QYE
4
Query Error. The power supply tried to read the output
buffer but it was empty. Or, new command line was
received before a previous query had been read. Or, both
the input and output buffers are full.
3
DDE
8
Device Error. A self-test or calibration error occurred
(see error numbers 601 through 750 in chapter 5).
4
EXE
16
Execution Error. An execution error occurred (see error
numbers -211 through -224 in chapter 5).
5
CME
32
Command Error. A command syntax error occurred (see
error numbers -101 through -178 in chapter 5).
6
Not Used
0
Always set to 0.
7
PON
128
Power On. Power has been turned off and on since the
last time the event register was read or cleared.
Summary of Contents for E364 A Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 27: ...Chapter 1 Quick Start To Rack Mount the Instrument 26...
Page 28: ...2 General Information...
Page 40: ...3 Front Panel Operation and Features...
Page 68: ...4 Remote Interface Reference...
Page 114: ...5 Error Messages...
Page 124: ...6 Application Programs...
Page 135: ...Chapter 6 Application Programs Example Program for Excel 97 134...
Page 136: ...7 Tutorial...
Page 147: ...Chapter 7 Tutorial Remote Programming 146...
Page 148: ...8 Specifications...
Page 156: ...Appendix Service Information...
Page 165: ...Appendix Service Information General Disassembly 164 General Disassembly...
Page 193: ...Appendix Service Information Replaceable Parts 192...
Page 197: ......
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......