Performance Test Record for Your Power Supply
CV Performance Test Record
Test Description
Models
Actual Result
Specifications
Upper Limit
Lower Limit
CV Programming Accuracy
@ 0 volts (DVM reading)
all
+0.0100 V
-0.0100 V
CV Readback Accuracy
@ 0 volts
all
DVM +0.0050 V
DVM -0.0050 V
CV Programming Accuracy
@ Full Scale (DVM reading)
E3640A/42A/44A
+20.0200 V
+19.9800 V
E3641A/43A/45A
+60.0400 V
+59.9600 V
CV Readback Accuracy
@ Full Scale
E3640A/42A/44A
DVM + 0.0150 V
DVM - 0.0150 V
E3641A/43A/45A
DVM + 0.0350 V
DVM - 0.0350 V
CV Load Effect (Load
Regulation)
E3640A/42A/44A
Maximum change: < 5 mV
E3641A/43A/45A
Maximum change: < 9 mV
CV Source Effect (Line
Regulation)
E3640A/42A/44A
Maximum change: < 5 mV
E3641A/43A/45A
Maximum change: < 9 mV
CV PARD (Normal mode)
E3640A/42A/44A
< 5 mVp-p / 0.5 mVrms
E3641A/43A/45A
< 8 mVp-p / 1 mVrms
Load Transient Response
Time
all
< 50
m
sec
Summary of Contents for E364 A Series
Page 9: ...8...
Page 15: ...14 Contents Contents...
Page 16: ...1 Quick Start...
Page 27: ...Chapter 1 Quick Start To Rack Mount the Instrument 26...
Page 28: ...2 General Information...
Page 40: ...3 Front Panel Operation and Features...
Page 68: ...4 Remote Interface Reference...
Page 114: ...5 Error Messages...
Page 124: ...6 Application Programs...
Page 135: ...Chapter 6 Application Programs Example Program for Excel 97 134...
Page 136: ...7 Tutorial...
Page 147: ...Chapter 7 Tutorial Remote Programming 146...
Page 148: ...8 Specifications...
Page 156: ...Appendix Service Information...
Page 165: ...Appendix Service Information General Disassembly 164 General Disassembly...
Page 193: ...Appendix Service Information Replaceable Parts 192...
Page 197: ......
Page 198: ......
Page 199: ......
Page 200: ......
Page 201: ......
Page 202: ......
Page 203: ......
Page 204: ......
Page 205: ......